Command Reference
R&S
®
ZNB/ZNBT
1128
User Manual 1173.9163.02 ─ 55
Example:
MMEM:LOAD:CKIT 'C:\Users\Public\
Documents\Rohde-Schwarz\Vna\
Calibration\Kits\New Kit 1 (123456).calkit'
Load the previously created cal kit file
New Kit 1 (123456).calkit
from the default cal kit direc-
tory.
CORR:CKIT:LSEL 'N 50 Ohm', 'New Kit 1',
'123456'
Assign the imported kit to the N 50 Ω connector type, assuming
that the cal kit name stored in
New Kit 1 (123456).calkit
reads
New Kit 1
and that its label reads
123456
.
[SENSe:]CORRection:CKIT:STANdard:LCATalog?
<CalKitName>, <KitLabel>
Returns a list of all standards in a given calibration kit.
Query parameters:
<CalKitName>
to query cal kit names and labels.
<KitLabel>
Label of the cal kit.
Example:
See
[SENSe<Ch>:]CORRection:CONNection:CATalog?
Usage:
Query only
[SENSe:]CORRection:CKIT:<ConnType>:LSELect
<CalKitName>, <Label>
Selects the calibration kit to be used for a specified connector type
<ConnType>
. The
kit is identified by its name and label.
Tip:
For connector types with arbitrary, user-defined names and labels you can use the
[SENSe:]CORRection:CKIT:LSELect
.
Suffix:
<ConnType>
.
Connector type, one of the following identifiers:
N50, N75
–
N 50 Ω or N 75 Ω connectors
PC7, PC35, PC292
–
PC 7, PC 3.5 or 2.92 mm connectors
USER<no>
–
user-defined connectors UserConn1, UserConn2
SMA
–
user-defined connector type SMA ...
Parameters:
<CalKitName>
String parameter containing the name of a calibration kit availa-
ble on the analyzer. See "Cal kit naming conventions " under
[SENSe:]CORRection:CKIT:<ConnType>:LSELect
*RST:
n/a (A *RST does not change the assignment
between connector types and calibration kits.)
<CalKitLabel>
String parameter containing the label of a calibration kit available
on the analyzer, usually the serial number.
*RST:
n/a (A *RST does not change the assignment
between connector types and calibration kits.)
SCPI Command Reference