Command reference
R&S
®
ZNB/ZNBT
1416
User Manual 1173.9163.02 ─ 62
SYSTem:ERRor:DISPlay:ERRor
<Boolean>
SYSTem:ERRor:DISPlay:INFO
<Boolean>
SYSTem:ERRor:DISPlay:WARNings
<Boolean>
Selectively disables or enables the display of information popups for the related event
type (Information, System Error, Warning).
Note
that the display of information popups can be globally disabled/enabled using
OFF/ON.
Parameters:
<Boolean>
ON | OFF - enable/disable display of information popups for the
related event type
*RST:
ON
Example:
SYST:ERR:DISP:STATe ON; ERR ON; WARN:OFF; INFO
OFF
Display information popups for system errors, but not for warn-
ings and information messages.
Manual operation:
See
"Show Info Messages / Show Warning Messages / Show
SYSTem:ERRor:DISPlay[:REMote]
<Boolean>
Disables or enables the display of information popups for remote control errors. These
popups appear at the bottom of the remote screen and the manual screen.
Note
●
Display of information popups can be globally disabled/enabled using
●
For SCPI error
-113, Undefined header
no tooltip is displayed
Parameters:
<Boolean>
ON | OFF - enable/disable display of information popups for
remote control errors
*RST:
ON
Example:
SYST:ERR:DISP:STATe ON; REM ON
Switch the display of information popups for remote command
errors on.
FREQ:STAR -1
Generate a
Remote Error: -222, "Data out of
range; ..."
tooltip.
Manual operation:
See
"Show Remote Error Info Messages"
SYSTem:ERRor:DISPlay:STATe
<Boolean>
Globally defines whether instrument events shall be indicated by information popups.
SCPI command reference