Basic operation
R&S
®
ZCxxx
26
Getting Started 1177.5156.02 ─ 05
6
Basic operation
This chapter describes how to configure the standard setup with an R&S
ZNA
network analyzer and two frequency converters R&S
ZC330, for 2-port transmis-
sion measurements.
Measurements using other converters of the R&S
ZCxxx family are performed in
an analogous way.
6.1
Required equipment
In general, the standard setup with
n
frequency converters requires an R&S
ZNA
with at least
n
test ports, equipped with the following hardware options:
●
Direct IF access (R&S
ZNA-B26)
●
Converter LO output (R&S
ZNA-B8)
The standard RF cabling and the required accessories are described in
To calibrate the measurement setup, the following equipment is required:
●
For power calibration and leveling, a power meter that covers the measured
frequency range:
– VDI Erickson PM5 or PM4 (with suitable taper) for waveguide ports
– Alternatively, R&S
NRP75TWG or R&S
NRP110TWG for waveguide ports
up to 110
GHz (scalar power calibration only)
– A supported coaxial power meter with an upper frequency limit of 23
GHz
or higher for calibrating the LO IN and RF IN signals at the converter (sca-
lar power calibration)
●
For system error correction, a calibration kit for the respective WM-xxxx wave-
guide
Required equipment