R&S UPZ
Manual Operation
Operating the R&S UPZ from the R&S UPV
Operating Manual 1120.8610.14 - 0
5
46
Fig. 2-20 Tracking selection: R&S UPV
indicates all possible settings in the R&S UPV control panel. The associated
remote control commands are listed. Further information on the syntax of the
R&S UPV remote control commands is given in the R&S UPV operating manual.
Table 2-2 List of settings and IEC/IEEE-bus commands: R&S UPV
Setting
Effect
IEC/IEEE-bus command
Switcher COM Port
Selection of the virtual COM Port on which
the R&S UPZ USB connection is installed.
SWITcher:COMPort COM<n>
SWITcher:COMPort AUTO
Switcher OFF
ON
R&S UPZ control is switched on
SWITcher:STATe ON|OFF
Input
A
n
Channel n of input switcher is switched to
busbar A
SWITcher:INPA <n>
Input
B
n
Channel n of input switcher is switched to
busbar B
SWITcher:INPB <n>
Output
A
n
Channel n of output switcher is switched to
busbar A
SWITcher:OUTA <n>
Output
B
n
Channel n of output switcher is switched to
busbar B
SWITcher:OUTB <n>
Tracking
OFF
B vs A
OUT vs INP
ALL
An additional selection window is opened
for channel tracking.
without tracking
only busbar B versus A
only output versus input
both tracking modes simultaneously
SWITcher:TRACking OFF
SWITcher:TRACking BVSA
SWITcher:TRACking OVSI
SWITcher:TRACking ALL
B vs A
n
The channel on busbar B is offset by <n>
versus A
SWITcher:OFFSet:BVSA <n>
Out vs Inp
n
The channel of the output switcher is offset
by <n> versus the input switcher
SWITcher:OFFSet:OVSI <n>