Startup
R&S
®
UPV
59
Getting Started 1146.2078.62 ─ 09
●
LAN LED off
–
The instrument does not have valid IP address. This may be due to one of the
following reasons:
1) the instrument is configured with DHCP, but no DHCP router is connected,
2) the instrument is manually configured and an IP address conflict is present,
3) the instrument lost its IP address that was configured with DHCP and
switched to Auto IP.
This state is called "error state" and normally can be resolved by the LCI mech-
anism (LAN reset). This mechanism is required by the LXI standard and can be
invoked with the "LAN RESET" button (key press > 10 s) on the rear side of the
instrument (see
chapter 2.2.11, "LAN RESET Pushbutton"
●
LAN LED is on
–
The instrument has a valid IP address and is reachable via TCP/IP.
●
LAN LED flashes
–
The instrument is in the "Device Indicator" state initiated by the WEB interface.
This state is required by the standard and is used for finding instruments in
racks (see
chapter 2.10.3.4, "LXI Browser Interface"
2.10.3.3
LXI Default State of Network Settings
After a LCI (LAN reset), the instrument automatically reboots and starts up with follow-
ing LAN settings:
Parameter
Setting
Hostname
RSUPV66vvv-xxxxxx
Description
Audio Analyzer
TCP/IP Mode
DHCP + Auto IP Address
Dynamic DNS
Enabled
ICMP Ping
Enabled
Negotiation
Auto Detect
VXI-11 Discovery
Enabled
Password for LAN Configuration
LxiWebIfc
Connect R&S UPV to a Network (LAN)