Instrument Function
R&S
®
SMJ100A
414
Operating Manual 1403.7458.32 ─ 14
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier:MODE
Number of Carriers
Sets the number of carriers for the multi carrier waveform.
By default the multi carrier table lists 1 carrier. When the number of carriers is
increased, the multi carrier table is extended by adding further lines at the end of the
table. If these carrier already have been defined before, the settings are preset accord-
ing to the former settings. Otherwise the parameters are preset with the default values.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier:COUNt
Carrier Spacing
Sets the frequency spacing between adjacent carriers of the multi carrier waveform.
The carriers are arranged symmetrically around the RF carrier and the maximum car-
rier spacing is limited to:
"Carrier Spacing" = Total Baseband Bandwidth /("Number of Carriers" - 1)
.
The total baseband bandwidth is 80 MHz.
Note:
In order to avoid wrap-around problems, the effective "Carrier Spacing" might be
slightly modified.
The "Carrier Spacing" is rounded in that way that the carrier closest to the center RF
frequency shows no phase jump assuming that the carrier is unmodulated.
●
For odd number of carriers:
RoundedCarrierSpacing=1/OutputSignalDuration* round(CarrierSpacing * Output-
SignalDuration);
●
For even number of carriers:
RoundedCarrierSpacing=2/OutputSignalDuration*round(0.5 *CarrierSpacing * Out-
putSignalDuration).
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier:SPACing
Crest Factor Mode
Selects the mode for optimizing the crest factor by calculating the carrier phases.
The crest factor represents the ratio of the peak voltage value to the rms voltage value.
The higher the crest factor and resulting dynamics of a signal, the greater the require-
ment for a power amplifier fed by the signal to be linear.
The following modes are available:
"Off"
There is no automatic setting for minimizing or maximizing the crest
factor. The "Phase" setting as defined in the carrier table is in use.
"Minimize"
The crest factor is minimized by internally calculating optimized car-
rier phases. The Phase setting displayed in the carrier table is invalid.
"Maximize"
The crest factor is maximized by internally calculating optimized car-
rier phases. The Phase setting displayed in the carrier table is invalid.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CFACtor:MODE
Baseband Signal - Baseband Block
Содержание SMJ100A
Страница 34: ...Preface R S SMJ100A 18 Operating Manual 1403 7458 32 14 Notes on Screenshots...
Страница 86: ...Preparing for Use R S SMJ100A 70 Operating Manual 1403 7458 32 14 LXI Configuration...
Страница 139: ...Manual Operation R S SMJ100A 123 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 140: ...Manual Operation R S SMJ100A 124 Operating Manual 1403 7458 32 14 Legend of Front Panel Controls...
Страница 454: ...Instrument Function R S SMJ100A 438 Operating Manual 1403 7458 32 14 Baseband Signal Baseband Block...
Страница 826: ...Remote Control Commands R S SMJ100A 810 Operating Manual 1403 7458 32 14 UNIT Subsystem...
Страница 828: ...Maintenance R S SMJ100A 812 Operating Manual 1403 7458 32 14 Storing and Packing...
Страница 843: ...Hardware Interfaces R S SMJ100A 827 Operating Manual 1403 7458 32 14 Figure A 2 BERT circuit diagram BERT Connector...
Страница 844: ...Hardware Interfaces R S SMJ100A 828 Operating Manual 1403 7458 32 14 BERT Connector...