DVB-T2 Configuration and Settings
R&S
®
SMCVB-K164
19
User Manual 1179.0974.02 ─ 02
IP TS Channel
Requires "Source > External" and "Input > IP".
Selects the IP-based transport stream (TS) channel. You can select 1 out of 4 IP TS
channels as input at the "IP Data" interface.
To configure a particular channel, see
Chapter 4.2.5, "IP Channnel x Settings"
Remote command:
[:SOURce<hw>]:BB:T2DVb:INPut:TSCHannel
T2-MI PID (Hex)
Requires "T2-MI Interface > On".
.
The PID belongs to MPEG transport stream packets, that contain T2-MI data.
Remote command:
[:SOURce<hw>]:BB:T2DVb:INPut:T2MI:PID
T2-MI SID (Hex)
Requires "T2-MI Interface > On".
.
Use the SID, when transmitting a composite signal, in accordance with annex
I of the
specification
Remote command:
[:SOURce<hw>]:BB:T2DVb:INPut:T2MI:PID
Stuffing
Requires "External/TS Player".
Activates stuffing.
"On"
Inserts null packets and corrects the
"Off"
The data rate of the transport stream source must match the data
rate required for the current modulation parameters.
Remote command:
[:SOURce<hw>]:BB:T2DVb:PLP<ch>:INPut:STUFfing
Test Signal
Requires "T2-MI Interface > Off" and "Source > Test Signal".
Defines the test signal data.
"Test TS Packet"
Standardized packet data used as modulation data in the transport
stream. To configure the packet structure, select the side tab "Test
Signal > Test TS Packet".
Uses a null packet as the test signal. To configure the null packet, see
Chapter 4.2.4, "Test Signal Settings"
Remote command:
[:SOURce<hw>]:BB:T2DVb:PLP<ch>:INPut:TESTsignal