Preparing for use
R&S
®
SMCV100B
28
Getting Started 1432.7046.02 ─ 04
2. Wait until the oven-controlled oscillator (OCXO) warms up. For the warm-up
time, see data sheet.
3. Press the [On/Standby] key.
Key and LED are located on the
.
The LED changes to green. The R&S
SMCV100B boots.
When starting for the first time, the R&S
SMCV100B starts with the default
settings. When restarting the instrument, the settings depend on the instru-
ment configuration before shut-down.
See the chapter "Saving and Recalling Instrument Settings" in the user man-
ual.
When the instrument is switched on, it automatically monitors main functions. You
can query erroneous functions. In addition to automatic monitoring, you can per-
form maintenance tasks.
See:
●
Chapter "Querying Error Messages" in the user manual.
●
Chapter "Performing Maintenance Tasks" in the user manual.
To shut down the product
The product is in the ready state.
► Press the [On/Standby] key.
The operating system shuts down. The LED changes to orange.
In the standby state, the power switch circuits and the OCXO are active. To deac-
tivate them, disconnect the instrument from the power supply.
To disconnect from power
The R&S
SMCV100B is in the standby state.
1.
NOTICE!
Risk of data loss. If you disconnect the product from power when it
is in the ready state, you can lose settings and data. Shut it down first.
Set the toggle switch on the power supply to position [0].
The LED of the [On/Standby] key is switched off.
2. Disconnect the R&S
SMCV100B from the power source.
Switching on or off