Reference Oscillator
R&S
®
SMA100B
182
User Manual 1178.3834.02 ─ 03
Remote command:
[:SOURce]:ROSCillator:INTernal:TUNing:SLOPe
7.4 Reference Output Settings
Access:
1. Select "Frequency" > "Reference Freq"
2. Select "Reference Output".
In the "Reference Output" tab, you can set the reference frequency value at the
output connectors.
The remote commands required to define these settings are described in
ter 12.15.11, "SOURce:ROSCillator Subsystem"
Settings:
Reference Output/1GHz Reference Output
................................................................ 182
Reference Output/1GHz Reference Output
Selects the signal that is to be output as frequency reference for downstream instru-
ments.
"Off"
Disables the reference signal output.
"10 MHz/100 MHz"
Derives a signal with 10 MHz or 100 MHz frequency from the internal
reference oscillator and forwards this signal to the output.
"1GHz"
Outputs a 1 GHz signal.
"Input Signal (loop through)"
Forwards the supplied reference frequency to the output directly.
Remote command:
[:SOURce]:ROSCillator:OUTPut:FREQuency:MODE
[:SOURce]:ROSCillator:OUTPut:ALTernate:FREQuency:MODE
Reference Output Settings