Remote Control Commands
R&S
®
SGT100A
331
User Manual 1176.8674.02 ─ 07
Example:
R&S
NRP-Z81
READ1?
Response:
-55.62403263352178,-22.419472478812476
-55.6 dBm is the measured average level, -22. 4 dBm is the
measured peak level at the given frequency.
Usage:
Query only
Manual operation:
See
" Level (Peak) / Level (Average) "
:SENSe<ch>:UNIT[:POWer]
<Power>
Selects the unit (Watt, dBm or dB
μ
V) of measurement result display, queried with
.
Parameters:
<Power>
DBM | DBUV | WATT
*RST:
DBM
Example:
SENS2:UNIT DBM
Selects dBm as unit for the measured value returned by com-
mand
READ
.
READ2?
Response: 7.34
7.34 dBm are measured by sensor 2.
Manual operation:
See
" Level (Peak) / Level (Average) "
:SENSe<ch>[:POWer]:APERture:DEFault:STATe
<UseDefAp>
Deactivates the default aperture time of the respective sensor.
To specify a user-defined value, use the command
Parameters:
<UseDefAp>
0 | 1 | OFF | ON
*RST:
1
Example:
SENS:POW:APER:DEF:STAT 0
Deactivates the default aperture time of the sensor.
Manual operation:
See
:SENSe<ch>[:POWer]:APERture:TIMe
<ApTime>
Defines the aperture time (size of the acquisition interval) for the corresponding sensor.
SENSe, READ, INITiate and SLISt Subsystems