![R&S RTO6 Скачать руководство пользователя страница 801](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646801.webp)
Protocol analysis
R&S
®
RTO6
801
User Manual 1801.6687.02 ─ 05
Remote command:
TRIGger<m>:TBTO:DADDress:CONDition
Source address
Sets the specified source address to be triggered on. The source address setup con-
sists of the condition and one or two data patterns.
The source address is the physical address of the device that sends the frame.
"Condition"
Defines the operator to set a specific source address, e.g. "Equal" or
"Not Equal") or a range.
"Data Min/Data"
Defines the bit pattern of the source address pattern.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Data Max"
The second data pattern is required to specify a range with conditions
"In range" and "Out of range".
Remote command:
TRIGger<m>:TBTO:SADDress:CONDition
Length/Type
Sets the specified length/type to be triggered on. The length/type setup consists of the
condition and one or two data patterns.
The value in this field indicates the manner in which the field is being used, either as a
length or type field. As a length field, the value in the field indicates the number of logi-
cal link control (LLC) data octets that follow in the data field of the frame. As a type
field, the value in the field is used to indicate the type of protocol data being carried in
the data field of the frame.
"Condition"
Defines the operator to set a specific Length/Type, e.g. "Equal" or
"Not Equal") or a range.
"Data Min/Data"
Defines the bit pattern of the Length/Type pattern.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Data Max"
The second data pattern is required to specify a range with conditions
"In range" and "Out of range".
Remote command:
TRIGger<m>:TBTO:LENGth:CONDition
Ethernet 1000BASE-T1 (option R&S
RTO6-K560)