Remote control commands
R&S
®
RTO6
2128
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..3
Parameters:
<DestAddrOptor>
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with
and
*RST:
EQUal
TRIGger<m>:TBTO:DADDress:MAX
<DestAddrPattMax>
Sets the end value of the destination address range if
is set to
INRange
or
OORange
.
Suffix:
<m>
.
1..3
Parameters:
<DestAddrPattMax> Numeric or string pattern, see
on page 1286. The string parameter accepts the bit
value X (do not care).
TRIGger<m>:TBTO:DADDress:MIN
<DestAddrPattMin>
Specifies the destination address bit pattern, or sets the start value of a pattern range.
Suffix:
<m>
.
1..3
Parameters:
<DestAddrPattMin>
Numeric or string pattern, see
on page 1286. The string parameter accepts the bit
value X (do not care).
TRIGger<m>:TBTO:DATA:DCONdition
<DataOperator>
Sets the condition for the data. You can define an exact value or a value range.