![R&S RTO6 Скачать руководство пользователя страница 663](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646663.webp)
Protocol analysis
R&S
®
RTO6
663
User Manual 1801.6687.02 ─ 05
Remote command:
Data Nibbles ← Transmission Sequence setup
Sets the number of data units in a single transmission sequence.
The maximum number of data nibbles is 6.
Remote command:
Data min ← Transmission Sequence setup
Sets the data pattern. Enter the bytes in msb first bit order. The maximum pattern
length is 64 bit. Waveform data is compared with the pattern byte-by-byte.
Remote command:
Data max ← Transmission Sequence setup
The second data pattern is required to specify a range with conditions "In range" and
"Out of range".
Remote command:
Serial Message setup
Configures the search conditions for a serial message.
Note:
The displayed parameters depend on the selected "Sequence". The instrument
indicates the data setting parameters when you select "ID+Data", see
on page 663.
The description of the search type specific settings are the same as for the trigger type,
see
Sequence ← Serial Message setup
Selects the condition for searching in a serial message.
You can search for the end of an identifier nibble or the combination of the identifier
and data nibble(s).
Remote command:
ID Type ← Serial Message setup
Selects the message ID format for the enhanced serial message type.
You can select either 4 bit or 8 bit message ID.
Remote command:
Identifier Condition ← Serial Message setup
Selects the operator to set a specific identifier or an identifier range.
The available operators:
●
Equal, Not equal
SENT (option R&S
RTO6-K520)