Protocol analysis
R&S
®
RTO6
527
User Manual 1801.6687.02 ─ 05
Pattern ← Data conditions
Specifies the data pattern to be found on the specified trigger source, in binary or hex
format. Enter the words in msb first bit order. The starting point of the pattern is defined
by
and
.
In binary format, use the following characters: 1; 0; or X (do not care). To enter the pat-
tern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
Remote command:
Position ← Data conditions
Operator for the data position. You can define an exact position, or a position range.
The setting is available if
is enabled in the protocol configuration.
Remote command:
Index min, Index max ← Data conditions
Sets the number of words before the first word of interest. These offset words are
ignored. If the
defines a range, the first and the last words of interest are
defined.
The setting is available if packet detection is enabled in the protocol configuration.
Remote command:
13.4.3.2
Triggering on UART signals
Prerequisites: An UART bus is configured, see
Chapter 13.4.3.1, "UART trigger set-
1. Press the [Bus] key and select the "Shortcuts" tab.
2. Press "Setup trigger".
3. Tap the "Source" button and select the "Serial bus" trigger source.
4. Select the serial bus that is set to UART.
5. Select the "Trigger type" and "Trigger source".
6. For more complex trigger types, enter the data conditions: pattern, position, index.
13.4.4
UART decode results
When the configuration of the serial bus is complete, the signal can be decoded:
1. In the "Serial Bus" dialog > "Setup" tab, enable "Decode".
UART/RS-232/RS-422/RS-485 (option R&S
RTO6-K510)