![R&S RTO6 Скачать руководство пользователя страница 338](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646338.webp)
Measurements
R&S
®
RTO6
338
User Manual 1801.6687.02 ─ 05
Remote command:
MEASurement<m>:SPECtrum:CPOWer:CFRequency
Occup. BW
Percentage of the total power used to determine the occupied bandwidth.
Remote command:
MEASurement<m>:SPECtrum:OBANdwidth
Threshold
Defines an absolute threshold as an additional condition for the peak search. Only
peaks that exceed the threshold are detected.
This setting is only available for spectrum waveforms. It is valid for cursor measure-
ments, spectrum measurements and peak search.
Remote command:
MEASurement<m>:SPECtrum:ATHReshold
Peak excursion
Defines a relative threshold, the minimum level value by which the waveform must rise
or fall to be considered as a peak. To avoid identifying noise peaks, enter a peak
excursion value that is higher than the noise levels.
This setting is only available for spectrum waveforms. It is valid for cursor measure-
ments, spectrum measurements and peak search.
The following figure shows a cursor measurement on a spectrum waveform:
Automatic measurements