Measurements
R&S
®
RTO6
317
User Manual 1801.6687.02 ─ 05
The R&S
RTO6 provides various voltage, time, area and counting measurements in
the category "Amp/Time". Some measurements require reference levels to be set
according to the measurement purpose.
Reference levels are explained in
Chapter 8.2.4, "Reference levels"
..................................................................................... 317
............................................................................................. 318
..............................................................................................320
Amplitude measurements
Access: [Measure] > "Setup" tab > "Amp/Time" category
Table 8-1: Amplitude measurements
Meas. type
Symbol
Description/Result
High
X
High
High signal level
Low
X
Low
Low signal level
Amplitude
X
Ampl
Amplitude of the signal: the difference of high and low signal lev-
els
X
Ampl
= X
High
- X
Low
Max
X
Max
Absolute maximum value of the waveform
Min
X
Min
Absolute minimum value of the waveform
Peak to peak
X
PkPk
Peak-to-peak value of the waveform: the difference of maximum
and minimum values
X
Ampl
= X
Max
- X
Min
Mean
X
Mean
Arithmetic average of the waveform voltage values
Eval
N
i
Eval
Mean
i
x
N
X
1
)
(
1
RMS
X
RMS
RMS (root mean square, quadratic mean) of the waveform volt-
age values
Eval
N
i
Eval
RMS
i
x
N
X
1
2
)
(
1
σ
σ
(S-dev)
σ
X
Standard deviation of the waveform samples
Eval
N
i
Mean
Eval
X
X
i
x
N
1
2
)
(
1
1