Remote control commands
R&S
®
RTO6
2518
User Manual 1801.6687.02 ─ 05
TRIGger<m>:LEVel<n>[:VALue]
<Level>
Sets the trigger level for the specified event and source.
If the trigger source is serial bus, the trigger level is set by the thresholds in the proto-
col configuration.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
<n>
1..31
Indicates the trigger source:
1...4 = channel 1 to 4
5 = external trigger input
6...11 = not available
12...27 = R&S
RT-ZVC input channels
28...31 = differential signals Diff1, Diff2 and Common1, Com-
mon2
Parameters:
<Level>
Voltage for the trigger level.
Range:
Depends on vertical scale, channel offset and other
settings. The trigger level must be within the current
display range.
Increment: 1E-3
*RST:
0
Default unit: V
Example:
TRIG:LEV5 0.01
Sets the trigger level for the external trigger signal to 10 mV.
TRIG2:LEV3 0.2
Sets the trigger level for the B-event and B-trigger source chan-
nel 3 to 200 mV.
Usage:
Asynchronous command
24.22.5
Serial pattern trigger using CDR
......................................................................................... 2518
........................................................................................... 2519
TRIGger<m>:SPATtern:CDR
<CdrInUse>
Disables the clock source and enables the recovered clock signal generated by hard-
ware CDR.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Jitter analysis and clock data recovery