User Manual 1801.6687.02 ─ 05
EQUal | LTHan | LETHan | GTHan | GETHan
Equal, Less than, Less or equal than, Greater Than, Greater or
equal than. These conditions require one pattern to be set using
INRange = RANGe
In range: Set the minimum and maximum value using
*RST:
INRange
TRIGger<m>:RFFE:DATA:IMIN
<DataIndexMin>
Specifies a pattern, or sets the start value of the index range.
Parameters:
<DataIndexMin>
Range:
1 to 32
Increment: 1
*RST:
1
TRIGger<m>:RFFE:DATA:IMAX
<DataIndexMax>
Sets the end value of a range, if the index condition is set to
INRange
or
OORange
.
See
TRIGger<m>:RFFE:DATA:ICONdition
.
Parameters:
<DataIndexMax>
Range:
1 to 32
Increment: 1
*RST:
32
TRIGger<m>:RFFE:SADD:CONDition
<SIDOperator>
Sets the operator to trigger on a specific slave address/MID pattern or a range.
Parameters:
<SIDOperator>
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater than,
Greater or equal than. These conditions require one pattern to
be set with
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with
and
*RST:
EQUal
Protocols