Remote control commands
R&S
®
RTO6
1988
User Manual 1801.6687.02 ─ 05
TRIGger<m>:ARINc:LABel:CONDition
<LabelOperator>
Sets the condition for the label. You can define an exact label or a label range.
Parameters:
<LabelOperator>
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with
.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with
and
*RST:
EQUal
TRIGger<m>:ARINc:LABel:MIN
<LabelMin>
Defines the minimum bit pattern for the label.
Parameters:
<LabelMin>
Numeric or string pattern, see
on page 1286. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:ARINc:LABel:MAX
<LabelMax>
Sets the end value of a label pattern if
TRIGger<m>:ARINc:LABel:CONDition
is
set to
INRange
or
OORange
.
Parameters:
<LabelMax>
Numeric or string pattern, see
TRIGger<m>:ARINc:MINGap:SELect
<MinGapSelect>
Enables triggering when the minimum gap is out of range. You can set the minimum
gap with:
.
Parameters:
<MinGapSelect>
ON | OFF
*RST:
ON
TRIGger<m>:ARINc:MINGap:BITS
<MinGapBits>
Sets the value for the minimum gap to be triggered on.
Protocols