Remote control commands
R&S
®
RTO6
1768
User Manual 1801.6687.02 ─ 05
24.17.7.2
Trigger
The trigger suffix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
To trigger on a serial bus, make sure that:
●
SBUS
.
●
The sources of the serial bus are channel signals: use
BUS<m>:...:SOURce
commands.
●
Decoding is enabled:
ON
.
............................................................................................1768
........................................................................................ 1769
.................................................................................... 1769
............................................................................................. 1769
............................................................................................ 1770
........................................................................... 1770
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:ENABle
....................................................... 1770
TRIGger<m>:CMSB:ADVanced:FRENable
..................................................................... 1770
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:ENABle
...........................................1770
TRIGger<m>:CMSB:ADVanced:FIENable
.......................................................................1770
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:BIT
................................................. 1771
............................................................................... 1771
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:DMAX
.............................................1771
...........................................................................1771
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:DMIN
..............................................1772
............................................................................1772
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:DOPerator
...................................... 1772
TRIGger<m>:CMSB:ADVanced:DOPerator
.....................................................................1772
TRIGger<m>:CMSB:ADVanced:ERRor<n>:ENABle
........................................................ 1772
TRIGger<m>:CMSB:ADVanced:ERENable
.....................................................................1772
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:IMAX
.............................................. 1773
............................................................................ 1773
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:IMIN
............................................... 1773
............................................................................. 1773
TRIGger<m>:CMSB:ADVanced:FRAMe<n>:FLD<o>:IOPerator
........................................ 1773
TRIGger<m>:CMSB:ADVanced:IOPerator
...................................................................... 1773
TRIGger<m>:CMSB:TYPE
<Type>
Selects the trigger type for custom serial bus analysis.
Suffix:
<m>
.
1..3
Parameters:
<Type>
STARt | PATTern | ADVanced
STARt
Triggers on the frame start, which is the end of the gap time as
specified in
Protocols