Remote control commands
R&S
®
RTO6
1681
User Manual 1801.6687.02 ─ 05
BIT7_RW
Enter 7 address bits and the R/W bit. Only available if
BIT10
10-bit address
ANY
Only available for trigger type "A data" (
). Used to trigger on data only, regardless of
the address.
*RST:
BIT7
Usage:
Asynchronous command
TRIGger<m>:I2C:ACONdition
<AddressOperator>
Sets the operator to set a specific address or an address range. The address values
are set with
.
Parameters:
<AddressOperator>
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
*RST:
EQUal
TRIGger<m>:I2C:ADDRess
<Address>
Triggers on the specified slave address, or sets the the start value of an address range
depending on the condition set with
.
Parameters:
<Address>
Numeric or string pattern, see
TRIGger<m>:I2C:ADDTo
<AddressTo>
Sets the the end value of an address range if the condition is set to an address range
with
Parameters:
<AddressTo>
Numeric or string pattern, see
Usage:
Asynchronous command
TRIGger<m>:I2C:ADOR<n>:ENABle
<UseAddress>
Includes the indicated ADOR address in the "address OR" trigger condition.
Suffix:
<n>
.
1..4
Index of the address in an "address OR" condition (OR slot)
Protocols