Remote control commands
R&S
®
RTO6
1363
User Manual 1801.6687.02 ─ 05
PROBe<m>:SETup:OFFSet:ZADJust
<ZroAdjVal>
set the waveform to zero position. It corrects the effect of a voltage offset or tempera-
ture drift. To set the value by the instrument, use
.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ZroAdjVal>
Range:
-100 to 100
Increment: 0.1
*RST:
0
Default unit: %
Usage:
Asynchronous command
Firmware/software:
FW 2.50
24.8.7.9
Probe attributes
..........................................................................................1363
...............................................................................................1363
.......................................................................................1364
..........................................................................................1364
.................................................................................1364
................................................................................... 1364
PROBe<m>:ID:SWVersion?
Queries the version of the probe firmware.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<Softwareversion>
Version number in a string.
Usage:
Query only
Asynchronous command
PROBe<m>:ID:PRDate?
Queries the production date of the probe.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<ProductionDate>
Date in a string.
Acquisition and setup