Remote control commands
R&S
®
RTO6
1326
User Manual 1801.6687.02 ─ 05
STOP
Stops the running acquistion.
Usage:
Event
Asynchronous command
24.8.2
Time base
.....................................................................................1327
........................................................................................1328
............................................................................................. 1329
...........................................................................................1329
..............................................................................................1329
.............................................................................................. 1330
.............................................................................................. 1331
...............................................................................................1331
................................................................................................1331
TIMebase:SCALe
<TimeScale>
Sets the horizontal scale - the time per division on the x-axis - for all channel and math
waveforms.
The setting accuracy depends on the current resolution (sample rate).
●
No interpolation:
The resolution is an integer multiple of the ADC sample rate.
●
With interpolation:
Any value for the horizontal scale can be set.
Parameters:
<TimeScale>
Range:
25E-12 to 10000 (RTO, RTP) | 5000 (RTE)
Increment: 1E-12
*RST:
10E-9
Default unit: s/div
Usage:
Asynchronous command
Acquisition and setup