Remote control commands
R&S
®
RTO6
1266
User Manual 1801.6687.02 ─ 05
24.3.4
Search
Searching for a pulse of specified width
..............................................................1266
24.3.4.1
Searching for a pulse of specified width
In the following example
*OPC
prevents overlapping execution of asynchronous com-
mands.
Search for positive pulses with pulse width 12 ± 10 μs (2 μs to 22 μs).
The usage of asynchronous commands is described in
, chapter "Command Sequence and Synchronization".
SEAR:ADD 'MySearch' // Create a new search
SEAR:TRIG:WIDT:STAT 'MySearch',1 // Configure search type
SEAR:SOUR 'MySearch',M1 // Configure search source - here Math1
SEAR:TRIG:WIDT:RANG 'MySearch',WITH // Configure search parameters
SEAR:TRIG:WIDT:WIDT 'MySearch',5e-4 // Configure search parameters
SEAR:TRIG:WIDT:DELT 'MySearch',5e-6 // Configure search parameters
SEAR:RES:LIM 'MySearch',1 // Set number of result lines in table to 1
*OPC?
SEAR:ALL 'MySearch'; // Initiate search for all events, asynchr. command
24.3.5
Data management
..................................................................................1266
Exporting waveform data to file
...........................................................................1267
Exporting measurement results to file
.................................................................1273
24.3.5.1
Saving a screenshot to file
Saves three display images in png format to the files
Print.png
,
Print_001.png
,
and
Print_002.png
in the directory
C:\Temp
. To get a correct screenshot, turn on
the display first.
Chapter 24.16.8, "Screenshots"
In the following example
*OPC
prevents overlapping execution of asynchronous com-
mands.
SYST:DISP:UPD ON
HCOP:DEST 'MMEM'
HCOP:DEV:LANG PNG
*OPC?
MMEM:NAME 'C:\Temp\Print.png'
HCOP:IMMediate; *OPC?
HCOP:IMM:NEXT; *OPC?
HCOP:IMM:NEXT; *OPC?
Programming examples