Instrument Tour
R&S
®
RTC1000
16
Getting Started 1335.7346.02 ─ 06
9 = Connector for the logic probe (option R&S
RTC-B1)
10 = Multi-purpose BNC connector [Aux Out]
11 = USB connector
4.1.1
Input Connectors
BNC inputs (6, 7)
The R&S
RTC1000 has two channel inputs (6) to connect the input signals. The
external trigger input (7) is used to control the measurement by an external sig-
nal. The trigger level can be set from -5
V to 5
V.
The input impedance of all BNC inputs is 1
M
Ω
.
Risk of electrical shock - maximum input voltages
The maximum input voltage on
channel inputs
must not exceed
200
V
(peak) and 150
V
(RMS).
For the
external trigger input
, the maximum input voltage is 100
V
(peak)
and 70
V
(RMS).
Transient overvoltages must not exceed 200
V
(peak).
Voltages higher than 30
V
(RMS) or 42
V
(peak) or 60
V
DC are regarded
as hazardous contact voltages. When working with hazardous contact vol-
tages, use appropriate protective measures to preclude direct contact with
the measurement setup:
●
Use only insulated voltage probes, test leads and adapters.
●
Do not touch voltages higher than 30
V
(RMS) or 42
V
(peak) or
60
V
DC.