Remote Commands Reference
R&S
®
RTC1000
270
User Manual 1335.7352.02 ─ 02
Suffix:
<b>
.
1, 2
Selects the bus.
Parameters:
<Source>
CH1 | CH2 | D0 .... D7 | EXTern
*RST:
EXT
BUS<b>:SPI:CS:POLarity
<Polarity>
Selects whether the chip select signal is high active (high = 1) or low active (low = 1).
Suffix:
<b>
.
1, 2
Selects the bus.
Parameters:
<Polarity>
POSitive | NEGative
POSitive
High active
NEGative
Low active
*RST:
POSitive
BUS<b>:SPI:CLOCk:SOURce
<Source>
Selects the input channel of the clock line.
Suffix:
<b>
.
1, 2
Selects the bus.
Parameters:
<Source>
CH1 | CH2 | D0 .... D7
*RST:
CH1
BUS<b>:SPI:CLOCk:POLarity
<Polarity>
Selects if data is stored with the rising or falling slope of the clock. The slope marks the
begin of a new bit.
Suffix:
<b>
.
1, 2
Selects the bus.
Protocol Analysis