CALCulate:MATH<m>:DATA:YINCrement?
●
CALCulate:MATH<m>:DATA:YORigin?
●
CALCulate:MATH<m>:DATA:YRESolution?
CALCulate:MATH<m>:DATA?
Returns the data of the math waveform points for transmission from the instrument to
the controlling computer. The waveforms data can be used in MATHLAB, for example.
Suffix:
<m>
.
1..5
Return values:
<Data>
List of values according to the format settings - voltages, or
magnitudes of a spectrum.
Usage:
Query only
CALCulate:MATH<m>:DATA:HEADer?
Returns information on the math waveform.
Table 17-3: Header data
Position
Meaning
Example
1
XStart in s
-9.477E-008 = - 94,77 ns
2
XStop in s
9.477E-008 = 94,77 ns
3
Record length of the waveform in Samples
120000
4
Number of values per sample interval, usually 1.
1
Suffix:
<m>
.
1..5
Return values:
<Header>
Comma-separated value list
Example:
-9.477E-008,9.477E-008,120000,1
Usage:
Query only
CALCulate:MATH<m>:DATA:POINts?
Returns the number of data samples that are returned with
.
Suffix:
<m>
.
1..5
Selects the math waveform.
Return values:
<DataPoints>
Number of data points
Documenting results