![R&S RTA4000 Скачать руководство пользователя страница 497](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987497.webp)
Remote control commands
R&S
®
RTA4000
497
User Manual 1335.7898.02 ─ 08
Parameters:
<SetupTime>
Range:
Depends on time base and sample interval
17.6.4.8
Pattern search configuration
................................................................................ 497
SEARch:TRIGger:PATTern:FUNCtion
...............................................................................497
SEARch:TRIGger:PATTern:LEVel<n>
...............................................................................498
SEARch:TRIGger:PATTern:LEVel<n>:DELTa
.................................................................... 498
SEARch:TRIGger:PATTern:WIDTh:RANGe
.......................................................................498
SEARch:TRIGger:PATTern:WIDTh[:WIDTh]
......................................................................499
SEARch:TRIGger:PATTern:WIDTh:DELTa
........................................................................ 499
SEARch:TRIGger:PATTern:SOURce
<Pattern>
Specifies the search pattern - the state for each channel. The state can be set only for
active channels.
Parameters:
<Pattern>
String parameter
String containing 0, 1, X|x for each channel. The order of chan-
nels is fixed: CH1 CH2 [CH3 CH4] D0 D1 D2 D3 ...
Example:
CHAN1:STAT ON
CHAN2:STAT ON
CHAN3:STAT ON
CHAN4:STAT ON
SEAR:STAT ON
SEAR:COND PATT
SEAR:TRIG:PATT:SOUR '1X10'
CH1, CH3 are high, CH4 is low. These states are logically com-
bined with
SEARch:TRIGger:PATTern:FUNCtion
not matter (don't care) and can be off.
SEARch:TRIGger:PATTern:FUNCtion
<Function>
Sets the logical combination of the channel states.
Parameters:
<Function>
AND | OR | NAND | NOR
AND
The required states of all channels must appear in the input sig-
nal at the same time.
OR
At least one of the channels must have the required state.
NAND
"Not and" operator, at least one of the channels does not have
the required state.
Waveform analysis