Remote control commands
R&S
®
RTA4000
492
User Manual 1335.7898.02 ─ 08
17.6.4.4
Peak search configuration
.................................................................................. 492
SEARch:MEASure:LEVel:PEAK:MAGNitude
.....................................................................492
SEARch:MEASure:PEAK:POLarity
<Polarity>
Indicates the polarity of a the pulse to be searched for a peak.
Parameters:
<Polarity>
POSitive | NEGative | EITHer
*RST:
POS
SEARch:MEASure:LEVel:PEAK:MAGNitude
<Magnitude>
Sets the peak-to-peak limit. If the signal exceeds this limit, a search event is listed.
Parameters:
<Magnitude>
Default unit: V
17.6.4.5
Rise/fall time search configuration
................................................................................. 492
SEARch:TRIGger:LEVel:RISetime:LOWer
........................................................................ 492
SEARch:TRIGger:LEVel:RISetime:UPPer
.........................................................................493
................................................................................ 493
....................................................................................493
.................................................................................. 493
SEARch:TRIGger:RISetime:SLOPe
<Polarity>
Sets the slope to be found.
Parameters:
<Polarity>
POSitive | NEGative | EITHer
POSitive: to search for rise time.
NEGative: to search for fall time.
EITHer: to search for rise and fall time.
*RST:
POS
SEARch:TRIGger:LEVel:RISetime:LOWer
<LowerLevel>
Sets the lower voltage threshold. When the signal crosses this level, the rise time mea-
surement starts or stops depending on the selected slope.
Parameters:
<LowerLevel>
*RST:
400 mV
Default unit: V
Waveform analysis