Remote control commands
R&S
®
RTA4000
471
User Manual 1335.7898.02 ─ 08
Parameters:
<PatternWidth>
Default unit: s
TRIGger:A:PATTern:WIDTh:DELTa
<PatternDelta>
Defines a range around the pattern width value specified using
.
Parameters:
<PatternDelta>
Default unit: s
17.5.7
Runt trigger
................................................................................. 471
..................................................................................471
.................................................................................................471
..............................................................................................472
............................................................................................... 472
................................................................................................ 472
TRIGger:A:LEVel<n>:RUNT:LOWer
<Level>
Sets the lower voltage threshold for runt detection. A positive runt crosses the lower
level twice without crossing the upper level.
Suffix:
<n>
.
1..5
Parameters:
<Level>
TRIGger:A:LEVel<n>:RUNT:UPPer
<Level>
Sets the upper voltage threshold for runt detection. A negative runt crosses the upper
level twice without crossing the lower level.
Suffix:
<n>
.
1..5
Parameters:
<Level>
TRIGger:A:RUNT:DELTa
<Variation>
Sets a range Δt to the reference pulse width set with
, if
is set to
WITHin
or
OUTSide
.
Parameters:
<Variation>
Range:
Depends on various settings, mainly time base and
sample rate
Trigger