![R&S RTA4000 Скачать руководство пользователя страница 385](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987385.webp)
Power analysis (option R&S
RTA-K31)
R&S
®
RTA4000
385
User Manual 1335.7898.02 ─ 08
Figure 14-14: Results of the SOA power measurement
The numerical results of the safe operating area mask test are described in
.
Table 14-6: Results of the mask test
Result
Description
Count
Number of tested points/acquisitions
Passed
Number of points/acquisitions that have passed the
mask test, i.e. they are within the defined safe opea-
rating area mask
Failed
Number of points/acquisitions that have failed the
mask test, i.e. they are out of the defined safe opea-
rating area mask
Fail rate
Ratio of acquisition hits to the number of tested
acquisitions
Result
A test has failed if the number of sample hits or
acquisition hits exceeds the limit of "Tot.Tolerance" /
"Acq. Tolerance" hits
Remote commands are described in
14.7.4.2
Configuring safe operating area measurements
For details of the configuration settings, see
Chapter 14.7.4.3, "Safe operating area
1. Press the Apps Selection key. Tap "Power Analysis".
2. In the "Power Path" tab, tap "SOA".
Power path power measurements