![R&S RTA4000 Скачать руководство пользователя страница 283](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987283.webp)
Serial bus analysis
R&S
®
RTA4000
283
User Manual 1335.7898.02 ─ 08
"Pattern"
Triggers on a data pattern at a specified position.
The pattern setup consists of the
"Any Symbol"
Triggers on a pattern that occurs in one symbol at any position in a
frame.
See
Remote command:
Symbol Offset
Sets the number of symbols at the beginning of the package to be ignored before the
serial pattern. The first symbol of interest is the first byte after the offset symbols.
The minimum offset is 0
symbols, the maximum offset is 4,095
symbols.
Remote command:
Numb. of Symb.
Sets the number of symbols (full bytes) you want to trigger on. The minimum is 1 sym-
bol, a maximum of 4 symbols is possible.
Note:
Entering data bits beyond the specified length of the pattern automatically
adjusts the "Number of Bytes" to include all specified bytes.
Remote command:
Data
Specifies the data pattern if "UART Trigger" is set to "Pattern" or "Any Symbol".
An example of pattern definition is shown in
To enter the binary value of any bit, tap this bit. To enter the hexadecimal value, tap
one of the nibbles (half byte) in the lower data line.
If a nibble (half byte) contains 1, 2 or 3 "X" bits (don't care), the nibble value is repre-
sented by the character "$". If all 4 bits of a nibble are "X", the nibble itself is "don't
care", represented by the character "X".
Remote command:
13.4.4
UART decode results
When the configuration of the serial bus is complete, the signal can be decoded:
1. In the "Bus" menu, enable "Decode".
2. In the "Display" menu, select the result display settings.
Chapter 13.1.2, "Displaying decode results"
3. In the "Bus Table" menu, enable the "Bus Table". Adjust the table settings.
Chapter 13.1.3, "Bus table: decode results"
UART / RS232 (option R&S
RTA-K2)