Functional Check
R&S
®
RT
‑
ZD10/20/30
56
User Manual 1410.4550.02 ─ 05
8
Functional Check
The functional check confirms the basic operation of the R&S
RT
‑
ZD10/20/30 dif-
ferential probe. The functional check is not suitable for verifying compliance with
the probe specifications.
1. Connect the R&S
RT
‑
ZD10/20/30 to a Rohde
&
Schwarz oscilloscope as
described in
Chapter 2.1, "Connecting the Probe to the Oscilloscope"
2. Using a short lead and a mini clip, connect one of the signal sockets to the
square wave output of the oscilloscope.
3. Using a short lead and a mini clip, connect the other signal socket to the
probe ground connector of the oscilloscope.
4. Press the [Preset] key and then the [Autoset] key on the oscilloscope.
A square wave with 1
V amplitude between 0
V and 1
V is displayed on the
display.
5. Reverse the pins at the probe and repeat
.