Safety information

R&S

®

RT

Z1M

5

Manual 1801.3142.02 ─ 05

1

 

Safety information

The product documentation helps you use the product safely and efficiently. Follow the
instructions provided here and in the following chapters.

Intended use

The R&S

 

RT

Z1M high-impedance buffer amplifier is designed for measurements on

circuits that are only indirectly connected to the mains or not connected at all. It is not
rated for any measurement category.

The product is intended for the development, production and verification of electronic
components and devices in industrial, administrative, and laboratory environments.
Use the product only for its designated purpose. Observe the operating conditions and
performance limits stated in the data sheet.

The buffer amplifier is designed for usage with oscilloscopes that have a
Rohde

 

&

 

Schwarz probe interface. Supported Rohde

 

&

 

Schwarz oscilloscopes are lis-

ted in the product's data sheet.

Where do I find safety information?

Safety information is part of the product documentation. It warns you of potential dan-
gers and gives instructions on how to prevent personal injury or damage caused by
dangerous situations.

Printed "Safety Instructions for Oscilloscopes and Accessories" in many languages
is delivered with your Rohde

 

&

 

Schwarz oscilloscope and probes.

The documentation of the oscilloscope and measurement equipment provides
safety instructions when you need to take care during setup or operation.

Operating site

Only use the product indoors, and keep it dry. The product casing is not waterproof.

You can operate the product up to the altitude specified in the data sheet. The lowest
specified altitude for a product of the measurement setup defines the altitude for the
complete setup.

The product is suitable for pollution degree 2 environments where nonconductive con-
tamination can occur. For more information on environmental conditions such as ambi-
ent temperature and humidity, see the data sheet.

Electrical safety

Take the following measures for your safety:

The buffer amplifier and the measurement instrument must be grounded. Ground-
ing is typically ensured through the grounding of the measurement instrument to
which the buffer amplifier is connected.

Always connect the buffer amplifier to the oscilloscope first before connecting a
probe to the buffer amplifier and to the circuit under test.

Содержание RT-Z1M

Страница 1: ...R S RT Z1M High impedance buffer amplifier Manual 1801314202 Version 05 B1OZ2...

Страница 2: ...Email info rohde schwarz com Internet www rohde schwarz com Subject to change data without tolerance limits is not binding R S is a registered trademark of Rohde Schwarz GmbH Co KG Trade names are tr...

Страница 3: ...ervice accessories 10 3 Putting into operation 11 3 1 Connecting the buffer amplifier to the oscilloscope 11 3 2 Identification of buffer amplifier and probe 12 3 3 Connecting the buffer amplifier to...

Страница 4: ...Contents R S RT Z1M 4 Manual 1801 3142 02 05 5 4 Calibration interval 18 5 5 Storage and transport 18 5 6 Disposal 18 6 Functional test 20...

Страница 5: ...on how to prevent personal injury or damage caused by dangerous situations Printed Safety Instructions for Oscilloscopes and Accessories in many languages is delivered with your Rohde Schwarz oscillo...

Страница 6: ...ntegrity The R S RT Z1M can withstand a moderate amount of physical and electrical stress To avoid damage treat the buffer amplifier with care Do not exceed the specified voltage limits Connect the R...

Страница 7: ...s e g passive probes current probes and active probes with Rohde Schwarz probe interface The R S RT Z1M provides an integrated R S ProbeMeter which is used for probes without an own R S ProbeMeter Usu...

Страница 8: ...e contents 1 Unpack the product carefully 2 Retain the original packing material Use it when transporting or shipping the prod uct later 3 Using the delivery notes check the equipment for completeness...

Страница 9: ...f the oscilloscope The interface provides also a USB interface and supply voltages 5 V USB 12 V high power It transmits analog signals and digital data simultane ously The output of the R S RT Z1M is...

Страница 10: ...e 2 2 Supplied accessories Item Quantity Description Order number R S RT ZP10 with probe accessories 1 10 1 passive probe 1409 7550 00 2 4 2 Service accessories To order accessories for servicing the...

Страница 11: ...h 50 input impe dance only Oscilloscopes that support a 1 M input impedance do not support the buf fer amplifier During usage the probe slightly heats up Warming is normal behavior and not a sign of m...

Страница 12: ...buffer amplifier to the DUT Observe the following step order when connecting the components of the measure ment setup Also read and observe the instructions in Chapter 1 Safety information on page 5...

Страница 13: ...of the channel to which the buffer amplifier is connected For details refer to the user manual of the oscilloscope Figure 3 2 Offset of R S RT Z1M with R S RT ZP10 signal range 1V Div 3 5 2 AC DC swi...

Страница 14: ...precision compared to the oscilloscope s DC accuracy The DC voltage is measured continuously and runs parallel to the time domain measurement of the oscilloscope When the R S ProbeMeter is active the...

Страница 15: ...The transfer functions of the gain or attenuation modes are shown in Figure 4 1 The bandwidth of a system specifies the maximum frequency at which a purely sinus oidal signal is still transferred at...

Страница 16: ...all 7 modes The rise time of a system is inversely proportional to the bandwidth The following approximation applies BW trise 4 0 The minimum measurable rise time of the whole setup also depends on th...

Страница 17: ...e casing is not waterproof Note Do not use cleaning agents Solvents thinners acetone acids and bases can damage the labeling or plastic parts 2 Dry the product completely before using it 5 2 Contactin...

Страница 18: ...z ser vice center see Chapter 5 3 Returning for servicing on page 18 5 5 Storage and transport Protect the product against dust Ensure that the environmental conditions e g tem perature range and clim...

Страница 19: ...Rohde Schwarz has developed a disposal concept for the eco friendly disposal or recycling of waste material As a manufacturer Rohde Schwarz completely fulfills its obligation to take back and dispose...

Страница 20: ...results are influenced by the used oscilloscope Figure 6 1 Functional test 1 Connect the R S RT Z1M to a Rohde Schwarz oscilloscope as described in Chapter 3 1 Connecting the buffer amplifier to the...

Отзывы: