Measurements and result displays
R&S
®
QAR50
36
User Manual 1179.4992.02 ─ 02
8
Measurements and result displays
The R&S
QAR50 provides several measurements and result displays that you can use
to evaluate the quality of a DUT.
Measurements usually consist of two steps, normalization and the actual measure-
ment.
1. Set up the normalization.
2. Perform the normalization.
For details about measurement normalization, see
3. Place the DUT between the clusters.
4. Start the measurement.
For details about measurement, see
5. Evaluate the measurement results using the result displays.
For details about the result displays, see
Chapter 8.3, "Result displays"
8.1
Normalization
Normalization is a separate measurement that provide a baseline to which other mea-
surements are compared.
Reflection and transmission measurements must be normalized separately by two sep-
arate measurements.
It is not necessary to normalize every measurement. We recommend normalizing the
measurement regularly. Refer to the datasheet for more information about recommen-
ded normalization intervals.
In addition, the software shows a message if normalization becomes necessary
because of temperature changes or other circumstances.
Normalizing reflection measurements
Normalizing the reflection measurement requires a clean, smooth and flat metallic
plate for a high reflection, for example aluminum. Such a metal plate represents a DUT
with 100
% reflectivity. Make sure to use a plate that has a size of at least 300
mm x
300
mm.
If you own the verification kit, you can use the verification plate labeled "E" to normal-
ize the measurement.
1. Place the plate between the clusters at the location where the DUT would be.
Normalization