Remote Control Commands
R&S
®
NRX
206
User Manual 1178.5566.02 ─ 07
Parameters:
<value>
Depends on the current output unit of the measured value.
Range:
1e-18 W to1e18 W; -150 DBM to 210 DBM; PCT:
1e-18 PCT to 1e22 PCT; -200 DB to 200 DB
DISPlay[:WINDow<Window>]:ANALog:UPPer
<value>
DISPlay[:WINDow<Window>]:METer:UPPer
<value>
Sets the upper limit value of the analog scale.
Suffix:
<Window>
.
1 to 4
Measurement pane
Parameters:
<value>
DISPlay[:WINDow<Window>]:METer:LOWer
[SENSe<Sensor>:]TRACe:OFFSet:TIME
<time>
Adds an offset to the beginning of the trace sequence. Thus, the trace in the result dis-
play is moved in positive or negative x-direction. If you measure with more than one
power sensor, you can use this offset to arrange the traces to each other. The start of
recording relative to the trigger event is set using
CHANnel<Channel>]:DELay[:VALue]
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<time>
Range:
-15.0 to 15.0
*RST:
0.0
Default unit: s
[SENSe<Sensor>:]TRACe:TIME
<time>
Sets the duration of the trace.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<time>
Range:
8.3e-9 to 30.0
*RST:
0.01
Default unit: s
Manual operation:
Measurement Settings and Results