![R&S NRX-B4 Скачать руководство пользователя страница 303](http://html1.mh-extra.com/html/rands/nrx-b4/nrx-b4_user-manual_3240273303.webp)
Remote Control Commands
R&S
®
NRX
303
User Manual 1178.5566.02 ─ 05
14.6.2
Sensor Ports
Further information:
●
Chapter 14.6.8.2, "NRT Correction Settings"
INPut<Sensor>:PORT:SOURce:AUTO
............................................................................. 303
INPut<Sensor>:PORT:SOURce[:VALue]
...........................................................................303
INPut<undef>:TRIGger:IMPedance
..................................................................................303
INPut<Sensor>:PORT:SOURce:AUTO
<auto>
Enables or disables the automatic assignment of the forward direction.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<auto>
ON
The direction in which the greater power flows is taken as the
forward direction.
OFF
If the forward and reverse power are almost equal, the orienta-
tion of the power sensor is defined using
*RST:
0
Example:
INP2:PORT:SOUR:AUTO OFF
INPut<Sensor>:PORT:SOURce[:VALue]
<val>
INPut<Sensor>:PORT:SOURce:AUTO
OFF
is set.
Defines the forward direction for the given input.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<val>
Range:
1 to 2
*RST:
1
INPut<undef>:TRIGger:IMPedance
<impedance>
Sets termination resistance of the external trigger input. Choose the setting that fits the
impedance of the trigger source to minimize reflections on the trigger signals.
Configuring Sensors