Remote Control Commands
R&S
®
NRX
263
User Manual 1178.5566.02 ─ 05
<DirectionalChannel> 1 to 2
1 = forward, 2 = reflection (reverse)
Parameters:
<value>
CALCulate<Measurement>:RELative<DirectionalChannel>[:MAGNitude]:AUTO
<state>
Sets the current measured value as reference value.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<DirectionalChannel> 1 to 2
1 = forward, 2 = reflection (reverse)
Parameters:
<state>
OFF | ONCE
*RST:
OFF
CALCulate<Measurement>:RELative<DirectionalChannel>[:STATe]
<state>
Enables or disables the relative measurement.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<DirectionalChannel> 1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<state>
OFF | ON | SET
SET
Uses the current measurement value as reference value and
enables the relative measurement.
*RST:
OFF
Manual operation:
See
[SENSe<Sensor>:]FUNCtion:CONCurrent
<concurrent>
Requires the sensor interface for R&S NRT (R&S NRX-B9).
Enables or disables the usage of several measurement functions simultaneously.
Measurement Settings and Results