Remote Control Commands
R&S
®
NRX
251
User Manual 1178.5566.02 ─ 05
CALCulate<Measurement>:STATistics:AWGN[:STATe]
...................................................... 252
CALCulate<Measurement>:STATistics:SAMPles[:MINimum]
.............................................. 252
CALCulate<Measurement>:STATistics:TGATe:SELection
...................................................253
[SENSe<Sensor>:]STATistics:AVERage?
Queries the average power value calculated during a statistics measurement.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Query only
[SENSe<Sensor>:]STATistics:OFFSet[:TIME]
<time>
Sets the start of the time interval relative to the (possibly delayed) trigger time.
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:
or
CALCulate<Measurement>:STATistics:APERture
interval in which the power for the statistical evaluation is measured.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<time>
Range:
0.0 to 10.0
*RST:
0.0
Default unit: s
[SENSe<Sensor>:]STATistics:PEAK?
Queries the peak power value calculated during a statistics measurement (CCDF or
PDF).
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Query only
CALCulate<Measurement>:STATistics:POWer:AVG:DATA?
Queries the average power value of the power in the time-defined window.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Measurement Settings and Results