Remote Control Commands
R&S
®
NRX
239
User Manual 1178.5566.02 ─ 05
Queries the time the pulse requires to transition from the pulse base level to the pulse
top level.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Usage:
Query only
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:POSitive:
OCCurrence?
Effective for pulse analysis measurements.
Queries the time when the signal passes through the medial reference power level with
rising edge, referenced to the delayed trigger event. Indicates the start point of the first
power pulse within the analysis window.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Usage:
Query only
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:POSitive:
OVERshoot?
Effective for pulse analysis measurements.
Queries the height of the local maximum before a falling edge, divided by the pulse
amplitude:
Positive overshoot = 100 % x
Max. power - pulse top power
Pulse amplitude
DISPlay[:WINDow<Window>]:TRACe:
.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Usage:
Query only
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:SPERiod?
Effective for pulse analysis measurements.
Sets the number of samples per second.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Measurement Settings and Results