Sensor Configuration
R&S
®
NRX
111
User Manual 1178.5566.02 ─ 05
Remote command:
MEMory:TABLe:FREQuency:POINts?
Export file name, Import file name ← Primary Sensor Offsets, Secondary Sensor
Offsets dialogs ← Frequency Dependent Offset
Edits the filename for export/import.
Export table to file, Import table from file ← Primary Sensor Offsets, Secondary
Sensor Offsets dialogs ← Frequency Dependent Offset
Exports or imports the table specified under
Export file name, Import file name
Gamma Correction
Enables or disables the gamma correction. See also
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:SGAMma:CORRection:
STATe
Gamma Phase
Available if
is enabled.
Sets the phase angle of the complex reflection coefficient of the source.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:SGAMma:PHASe
Gamma Magnitude
Available if
is enabled.
Sets the magnitude of the complex reflection coefficient of the source.
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:SGAMma[:MAGNitude]
9.3
Filter Settings
Access:
"Measurement Settings" > "Primary Sensor Config" > "Filter"
"Measurement Settings" > "Secondary Sensor Config" > "Filter"
Available for all measurement types.
Use the averaging filter to reduce fluctuations in the measurement results. Such fluctu-
ations can be caused by inherent noise of the power sensor, modulation of the mea-
surement signal or beats from the superposition of adjacent carriers. A more stable dis-
play is traded off against longer measurement times, caused by longer settling times
Filter Settings