Browser-Based User Interface
R&S
®
NRPxxS(N)
52
User Manual 1177.5079.02 ─ 10
............................................................................................... 53
Averaging Mode
Groups the averaging settings. See also
Chapter 9.8.1, "Configuring Auto Averaging"
<Mode> ← Averaging Mode
Sets the averaging mode.
"manual"
Disables auto averaging. Enter the average count under
"auto 1
dB" / "auto 0.1
dB" / "auto 0.01
dB" / "auto 0.001
dB"
Uses an automatic averaging filter with the respective resolution
index.
"noise content" Predefines the compliance to an exactly defined noise component.
Enter this value under
Remote command:
[SENSe<Sensor>:]AVERage[:STATe]
[SENSe<Sensor>:]AVERage:COUNt:AUTO
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE
<Value> ← Averaging Mode
The content of this field depends on the setting under
●
If "manual" is set:
Sets the average count, also called averaging factor.
●
If "auto xx
dB" is set:
Displays the resolution index.
●
If "noise content" is set:
Sets the maximum noise ratio in the measurement result.
Remote command:
[SENSe<Sensor>:]AVERage:COUNt:AUTO:RESolution
[SENSe<Sensor>:]AVERage:COUNt:AUTO:NSRatio
Filter Terminal Control
Defines how the measurement results are output. This is called termination control.
See also
Chapter 9.5, "Controlling the Measurement"
"Repeating"
Outputs intermediate values to facilitate early detection of changes in
the measured quantity. In the settled state, that means when the
number of measurements specified by the averaging factor has been
performed, a moving average is output.
Settings