Getting Started
R&S
®
NRP2
29
User Manual 1173.9157.02 ─ 03
2.4 Measuring average power (Cont Av mode)
In the following example for measuring the average power the following steps are per-
formed:
●
Setting up the measurement
●
Setting the frequency
●
Setting the unit for the measurement result
●
Setting a fixed offset correction
●
Performing a relative power measurement
●
Checking the settings with full height window
●
Selecting the display mode
●
Selecting the indication of auxiliary values
Setting up the measurement
To set up the measurement
1. Put the R&S
NRP2 into operation as described in
chapter 1, "Putting into Opera-
, on page 7 and connect a sensor to connector A.
2. Preset the R&S
NRP2 and execute Zero
A.
After preset, one result window indicates the result (in dBm) obtained with sensor
A.
Fig. 2-3: Result window (dBm)
Setting the frequency
The R&S
NRP2 must be set to the carrier frequency of the applied signal if the specified
measurement accuracy is to be reached.
To set the frequency
1. Connect an unmodulated signal with a level between –10
dBm and +10
dBm to the
sensor.
2. Press the FREQ hardkey.
Measuring average power (Cont Av mode)
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...