Remote Control - Commands
R&S
®
NRP2
309
User Manual 1173.9157.02 ─ 03
exceeded since the last readout of the Operation Upper Limit Fail Status Register. Details
of the behavior are defined by the transition register. The limit value can be set with the
command
CALC:LIM:UPP:DATA <float_value>
. Command
CALC:LIM:UPP:STAT ON
configures the PTRansition and NTRansition registers so that
the corresponding bit is set in the EVENt register when the displayed value exceeds the
upper limit value.
The Operation Upper Limit Fail Status Register can be read by the commands:
STATus:OPERation:ULIMit[:SUMMary]:CONDition?
STATus:OPERation:ULIMit[:SUMMary][:EVENt]?
.
Table 6-28: Meaning of bits used in the Operation Upper Limit Fail Status Register
Bit No.
Meaning
0
Not used
1
Window 1 Upper Limit Fail
The measured value returned by the first Calculate block exceeds the upper limit value.
2
Window 2 Upper Limit Fail
The measured value returned by the second Calculate block exceeds the upper limit value.
3
Window 3 Upper Limit Fail
The measured value returned by the third Calculate block exceeds the upper limit value.
4
Window 4 Upper Limit Fail
The measured value returned by the fourth Calculate block exceeds the upper limit value.
5
Window 5 Upper Limit Fail
The measured value returned by the fifth Calculate block exceeds the upper limit value.
6
Window 6 Upper Limit Fail
The measured value supplied by the sixth Calculate block exceeds the upper limit value.
7
Window 7 Upper Limit Fail
The measured value supplied by the seventh Calculate block exceeds the upper limit value.
8
Window 8 Upper Limit Fail
The measured value supplied by the eighth Calculate block exceeds the upper limit value.
9 to14
Not used
15
Bit 15 will never be used.
6.13.17 Questionable Power Status Register
The CONDition register contains information as to whether the measured power values
are questionable.
The Questionable Power Status Register can be read by the following commands:
STATus:QUEStionable:POWer[:SUMMary]:CONDition?
STATus:QUEStionable:POWer[:SUMMary][:EVENt]?
STATus
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...