Remote Control - Commands
R&S
®
NRP2
251
User Manual 1173.9157.02 ─ 03
MEMory:TABLe:DATA:POINts?
The command returns the number of stored entries in the currently selected offset table.
Usage:
Query only
MEMory:TABLe:FREQuency
<NRf> {, <NRf>}
This command enters frequencies in the first column of the table selected with
MEM:TABL:SEL
. Existing data for frequencies will be overwritten. The number of fre-
quencies should match the number of offset values. If the number of frequencies differs
from the number of offset values, the shorter column length will be used.
SENS<[1]..4>:FREQ
can be used to determine a frequency for which a correction fac-
tor is then defined by means of the offset table. If the exact frequency selected is not in
the table, the values in the table are interpolated. If the selected frequency is outside the
specified frequency range, the first or last offset value of the table is selected. Interpola-
tion is linear in the units HZ and DB.
Parameters:
<NRf> {, <NRf>}
Enters frequencies in the first column of the table selected with
MEM:TABL:SEL
*RST:
Tables are not changed in case of a *RST.
Default unit: HZ
MEMory:TABLe:FREQuency:POINts?
Returns the number of entries (lines) in the table selected with
MEM:TABL:SEL
<name>
.
Usage:
Query only
MEMory:TABLe:GAIN[:MAGNitude]
<NRf> {, <NRf>}
This command enters offset values in the second column of the table selected with
MEM:TABL:SEL
. Existing data for offset values will be overwritten. The number of fre-
quencies should match with the number of offset values. If the number of frequencies
differs from the number of offset values, the shorter column length will be used.
A frequency can be determined with
SENS<[1]..4>:FREQ
, for which a correction factor
is then calculated by means of the offset table. If the exact frequency selected is not in
the table, the values in the table are interpolated. If the selected frequency is outside the
specified frequency range, the first or last offset value of the table is selected. Interpola-
tion is linear in the units HZ and DB.
Unit:
DB | PCT
MEMory
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...