Remote Control - Commands
R&S
®
NRP2
241
User Manual 1173.9157.02 ─ 03
Parameters:
<display_mode>
DIGital | ANALog | GRAPhical
DIGital
Measured values are displayed in numeric format. If the primary
channel is operated in Trace mode, the (scalar) measured values
of the currently selected timegate are displayed here.
ANALog
Measured values are indicated by a marker on a scale. The values
at the upper and lower scale end as well as the type of scaling
(linear or logarithmic) are set with the following
:METer
com-
mands.
GRAPhical
Measured values are plotted over time. This kind of presentation
is possible only if the primary channel is operated in Trace mode.
*RST:
DIGital
DISPlay[:WINDow]<[1]..4>:METer:AUTO
<auto>
DISPlay[:WINDow]<[1]..4>:ANALog:AUTO
<auto>
DISP:AUTO ONCE
automatically determines scaling for the analog display. The upper
and the lower limit value of the display is set as a function of the current measurement
data. Only the
ONCE
parameter is permissible; a query always supplies the value
OFF
.
Parameters:
<auto>
ONCE | OFF
*RST:
OFF
DISPlay[:WINDow]<[1]..4>:METer:LOWer
<lower_limit>
DISPlay[:WINDow]<[1]..4>:ANALog:LOWer
<lower_limit>
Sets the lower limit value of the analog scale.
The value range depends on the current output unit of the measured value of the window.
Table 6-12: Value range
Unit of measured value
W
DBM
PCT
DB
Value range for <float_value>
1e-18..1e18
-150..210
1e-18..1e22
-200..200
Unit:
Value ranges, depending on the unit
.
Parameters:
<lower_limit>
float_value
*RST:
-90 DBM or -120 DB
DISPlay
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...