Remote Control - Commands
R&S
®
NRP2
239
User Manual 1173.9157.02 ─ 03
DISPlay:PIXMap?
Supplies the display content as binary block data. The block data consists of a block data
header (#577110) and the display content in Microsoft Windows (R) BMP format (version
3). A color depth of 8 bit per pixel is used.
Parameters:
<pixmap>
block_data
Example:
DISP:PIXM?
returns:
#577110xxxxxx...x,
with x in binary format comprising an 8-bit BMP bitmap of the dis-
play content.
Usage:
Query only
DISPlay:UPDate
<update_frequency>
Adjusts the update frequency of the measured values in the display. In case of disconti-
nuities in the voltage progress at the analog outputs display updates may be suspended
with
DISP:UPD FREeze
. The display won't consume CPU time in this state.
Parameters:
<update_frequency> NORMal | SLOW | FREeze
*RST:
NORMal
DISPlay[:WINDow]<[1]..4>:AVALue
<avalue>
Determines which additional information about the measured values is shown in the dis-
play (right-hand side, next to the main measured values).
DISPlay
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...