Remote Control Fundamentals
R&S
®
NRP2
177
User Manual 1173.9157.02 ─ 03
Processing of a program message is aborted if an error occurs. If only sensor 2 is con-
nected to a multichannel device, for example, the program message
SENS1:FREQ 50 GHZ;:SENS2:FREQ 50 GHZ
will be aborted after the first command and a second frequency setting will not be per-
formed.
5.7.4 Responses to Queries
Unless otherwise expressly specified, a query is defined for each command. The query
is created by appending a question mark to the associated command.
Some of the SCPI rules for query responses are stricter than those of the IEEE 488.2
standard:
●
The required value is sent without header.
Example:
SENSe:AVERage:TCONTrol?
Response:
MOV
●
Maximum and minimum values as well as all other quantities queried by a special
text parameter are returned as numeric values
Example:
SENSe:POWer:TSLot:COUNt? MAXimum
Response:
26
●
Numeric values are output without a unit. Physical quantities refer to the basic units
or to the units set with the UNIT command.
Example:
SENSe3:FREQuency?
Response:
1.000000E06
(for 1 MHz)
●
Boolean values are returned as 0 (for OFF) and 1 (for ON).
Example:
SYSTem:BEEPer:STATe?
Response:
1
●
Character data is returned in short form (
Example:
UNIT2:POWer?
Response:
DBM
5.7.5 Parameters
Most commands require the specification of a parameter. Parameters must be separated
from the header by a white space. Parameters may be specified as numeric values,
Boolean parameters, character data, strings or block data. The type of parameter
required for the specific command as well as the permissible value range are described
together with the commands.
Structure and Syntax of device-dependent Messages
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...