Instrument Functions
R&S
®
NRP2
162
User Manual 1173.9157.02 ─ 03
Fig. 4-96: Sensor test results message
... with pad
Performs a test measurement for a sensor with an attached attenuator. The padded sen-
sor must be connected to the sensor check source. With multipath sensors, only one path
can be used for the measurement. For testing all paths, remove the attenuator, connect
the sensor section alone and start the test by pressing "Test at Power Ref".
Report…
Opens the dialog box displaying the results of the test measurement.
The test report shows the type and the serial number of the tested sensor and the detailled
test results.
When the dialog box opens, it shows the data of the sensor last tested. The measured
values are lost when the R&S
NRP2 is switched off.
Fig. 4-97: Sensor test report, page 1
Page 1 shows the test results for the sensor section.
Fig. 4-98: Sensor test report, page 2
Page 2 shows the results with attached attenuator, if applicable.
Differences up to 0.02 dB at room temperature are absolutely normal and are due to the
uncertainty in the calibration of the sensor and the power reference. Greater differences
can be caused by high/low ambient temperatures but also by a damage to the sensor.
System settings
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...