Instrument Functions
R&S
®
NRP2
152
User Manual 1173.9157.02 ─ 03
Fig. 4-83: Out1 mode / Pass/Fail
SCPI command:
Window 1 2 3 4 ← Pass/Fail ← Mode
Selects the window the output refers to.
Only windows that are operated in "Dig" or "D&A" window mode are available for selec-
tion. See
chapter 4.3.3.3, "Selecting the result display mode"
, on page 93 for the setting
the display mode.
SCPI command:
Pass ← Pass/Fail ← Mode
Sets the logic output level for measurement results within the set limits. The value range
is 0
V to 3.3
V.
SCPI command:
The setting is achieved by a combination of the following commands. The example is for
OUTPut:TTL:FAIL
set to
HIGH
.
Fail ← Pass/Fail ← Mode
Sets the logic output level for limit violation of measurement results. The value range is
0
V to 3.3
V.
SCPI command:
The setting is achieved by a combination of the following commands. The example is for
OUTPut:TTL:FAIL
set to
HIGH
.
TrigOut ← Mode
The output provides a logic signal, which represents the state of the internal trigger bus.
System settings
Содержание NRP2
Страница 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Страница 20: ......
Страница 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Страница 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Страница 66: ...Getting Started R S NRP2 48 User Manual 1173 9157 02 03 Measuring average burst power Burst Av mode...
Страница 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Страница 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Страница 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Страница 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Страница 418: ...Programming Measurements R S NRP2 400 User Manual 1173 9157 02 03 Optimization of measurement speed...
Страница 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...